Kaedah ujian dan langkah berjaga-jaga penguji kehilangan dielektrik minyak penebat

Feb 23, 2022

1. Uji voltan

Voltan AC bergantung pada cecair yang akan diukur, kekuatan medan elektrik adalah antara 0.03 kV/mm dan 1 kV/mm, dan voltan sinusoidal dengan frekuensi antara 40 dan 62 Hz digunakan.

2. Pengukuran

After the oil cup is filled with the oil sample, when the difference between the temperature of the inner electrode and the required test temperature is not greater than ±1 degree , the measurement of the loss factor starts, and the voltage is only applied during the measurement. After completing the first measurement, pour out the oil sample and refill the oil cup with a second sample of the same oil. The procedure is the same as the first time, but the rinsing is omitted. Repeat the measurement after warming, and the difference between the two readings should not be greater than 0.010 percent plus 25 percent of the larger value of the two measurements.

Jika keperluan ini tidak dapat dipenuhi, tukar sampel minyak yang sama sekali lagi dan lakukan ujian sehingga perbezaan antara dua bacaan bersebelahan tidak melebihi jumlah 0.010 peratus dan 25 peratus daripada nilai yang lebih besar bagi kedua-dua ukuran. . Pada masa ini, keputusan yang diukur dianggap sah. .

Akhir sekali, ambil bahagian bawah daripada dua ukuran yang sah sebagai faktor kehilangan bagi sampel cecair.


Kesalahan dan penyelesaian biasa

1. Apabila mesin dihidupkan, lampu penunjuk suis kuasa tidak menyala. Sila periksa teras fius papan kuasa untuk melihat sama ada ia terputus.

2. When the device is boosting the voltage, the LCD displays "electrode cup short circuit", please check whether the electrode cup is properly assembled.

3. When the device detects that the capacitance value of the empty cup deviates greatly from the standard value (60pF±5pF), please check whether the radio frequency head on the protective electrode cover of the power signal cable is loose.

4. Apabila peranti menjadi panas, isyarat suhu tidak dapat dikesan. Sila semak sama ada kabel isyarat suhu disambungkan dengan betul.

5. Apabila peralatan tidak panas (iaitu, tiada bunyi khas pemanasan frekuensi pertengahan), sila periksa sama ada insurans pemanasan ditiup.